Successful introduction of new anticancer agents into the clinic is often hampered by a lack of qualified biomarkers. Studies have been conducted of 17 ELISAs representing a potential panel of pharmacodynamic/predictive biomarkers for drugs targeted to tumour vasculature.
The fit-for-purpose approach to method validation was used. Stability studies were performed using recombinant proteins in surrogate matrices, endogenous analytes in healthy volunteer and cancer patient plasma. The impact of platelet depletion was investigated.
Method validation focused on measuring precision and showed that 15 of the 17 assays were within acceptable limits. Stability at −80°C was shown for 3 months with all recombinant proteins in surrogate matrices, whereas under the same conditions instability was observed with KGF in platelet-rich and platelet-depleted plasma, and with PDGF-BB in platelet-depleted plasma from cancer patients. For measurement of extracellular circulating analytes, platelet depletion should be conducted before freezing of plasma to prevent release of PDGF-BB, FGFb and VEGF-A. A protocol was developed to remove >90% platelets from plasma requiring centrifugation at 2000
These studies highlight the need for assay validation and crucial assessment of sample handling issues before commencement of biomarker analysis in clinical trials.
Angiogenesis, the formation of new blood vessels from existing vasculature, is required for tumour growth (
Judicious implementation of multi-modality biomarkers (tissue, imaging and blood-borne) could potentially enrich selection of likely responders and allow real-time monitoring of drug effects (
Studies using circulating biomarkers of angiogenesis have produced conflicting and often confusing results and this may reflect complex biology, differences in antibody versus small-molecule drugs and/or lack of assay validation (
Method validation constitutes a crucial component in biomarker research, and it is often the case that a biomarker can fail in the clinic not because of the underlying scientific rationale but rather from poor assay choice and lack of robust validation (
Seventeen ELISA kits (Quantikine Human - Sandwich ELISA – Immunoassay; R&D Systems Europe Ltd, Abingdon, UK) representing a comprehensive panel of putative biomarkers of angiogenesis were evaluated (see
Mindful of minimising patient sample volume, the Searchlight Multiplex ELISA system was also included (Aushon BioSystems, Boston, MA, USA) allowing further validation and a more comprehensive evaluation of sample stability. Two multiplex ELISAs were used: a 5-plex comprising VEGFR1, VEGFR2, IL8, keratinocyte growth factor (KGF) and PIGF, and a 4-plex comprising platelet-derived growth factor (PDGF-BB), HGF, FGFb and VEGF-A. Plex expiry dates were logged and plexes stored at 2–8°C before use. Both assays were used according to the manufacturer's instructions and had been subjected to extensive method validation by our laboratory before this study (
The performance of the 17 different single-plex ELISA assays in terms of precision was determined using quality-control (QC) samples at three different concentrations corresponding to the low end, middle and top end of the calibration curve for each analyte as detailed in
Throughout the duration of the studies, validation experiments were performed upon introduction of either a new batch of kit(s) or a new batch of QCs (
Lyophilised carrier-free recombinant (r) protein standards were stored at −20°C before reconstitution in a surrogate matrix for the stability studies. The reconstituted r-proteins were spiked at a mid-range concentration into porcine/bovine plasma and serum stocks (Scipac, Sittingbourne, UK) and 300-
Normal human plasma (prepared in EDTA) from two healthy volunteers was obtained from Scipac Ltd. Upon receipt the plasma samples were analysed by multiplex ELISA using eight replicates; they were then stored in aliquots at −80 °C for 8 months before re-analysis by multiplex ELISA (
All studies using patient samples were performed under ethical committee approval and all patients provided informed consent (REC Ref: 06/Q1406/117). Plasma samples for stability studies were obtained from five patients with colorectal cancer. Four aliquots were prepared from each patient sample by different centrifugation protocols, including platelet-depleted plasma (see below); they were analysed in triplicate upon receipt by multiplex ELISA and stored at −80°C for 3 months before re-analysis by multiplex ELISA. Instability was defined as a significant change in concentration (⩾25%) between the two time points.
To evaluate the impact of platelets on the measurement of the panel of angiogenesis-related analytes by ELISA, blood was collected from 20 patients with metastatic colorectal cancer who were receiving conventional chemotherapy at The Christie NHS Foundation Trust (Manchester, UK). A 20-ml volume of venous blood was withdrawn from each subject and transferred to an EDTA vacutainer and labelled as the whole-blood (WB) specimen. Aliquots of the WB specimen were retained for platelet count by the Haematology Department of The Christie. The WB specimen was centrifuged at 2000
The process developed to validate the panel of 17 ELISAs using QCs is described schematically in
Results on the above validation process for the 17 ELISAs are reported in
Stability studies were conducted in three stages: first by adding a known concentration of a r-protein to a surrogate matrix; second by measuring endogenous angiogenesis analytes in plasma collected from healthy volunteers and third by re-analysis of plasma samples taken from cancer patients. The stability of r-proteins spiked into plasma and serum, and stored at RT, 4°C, −20°C and −80°C, is reported in
Endogenous analytes were measured in pooled healthy volunteer plasma (
To evaluate stability in cancer patients' plasma, angiogenesis analytes were measured in four replicate plasma samples collected from five different patients: a study design that allowed between-patient comparison to be made in platelet-rich (
After 3 months, a significant change (increase) in concentration was detected consistently in all five patients with only one analyte in both platelet-rich (KGF,
Cancer patients' plasma was centrifugated in stages to determine the minimum duration and optimal speed to remove platelets effectively from plasma. The protocols adopted in this study yielded four different plasma samples: each was subjected to a standard procedure, but three samples received an additional spin of either increasing duration or centrifugal force (see Methods and
Of the analytes investigated, removal of platelets reduced significantly the plasma concentrations of PDGF-BB (mean reduction in five different patients of 77%), FGFb (63%) and VEGF-A (43%) (
The data show that removal of platelets reduced the plasma concentration of certain angiogenesis-associated factors, if the platelets were removed before freezing plasma samples. Once plasma samples were frozen and platelets presumably ruptured, then centrifugation was without effect on angiogenesis analyte concentrations (see
With the eventual objective of qualifying biomarkers to facilitate the clinical development of drugs targeted to the tumour vasculature, a panel of ELISAs for circulating angiogenesis associated factors was validated (see
The strategy adopted (
The success of this validation approach relies heavily on an accurate determination of the total error associated with each assay. Total error for an LBA is assumed to follow a normal distribution and consist of a systematic component (bias, measured as percent relative error (%RE)) and a random component (precision, measured as the coefficient of variation (%CV)) (
Evaluation of error and choice of acceptance limits was based on the results from 2–3 assays of eight replicate measurements of the QCs per run, which tended to weight the validation towards within-day/intra-assay precision. As this performance parameter often shows less variability than between-day/inter-assay precision, it is possible that the true level of imprecision was undervalued. Nonetheless, 15 of 17 assays did pass this stricter validation regime. However, an assessment of the errors associated with accuracy and bias, the systematic component in the total error model, was confounded by the fact that commercially available QC standards were used. These were provided by the manufacturer not at a nominal concentration together with a certificate of analysis but at high, medium and low concentration ranges, and thus could not be added at known concentrations. The issue of poorly characterised, or non-representative (recombinant proteins or peptide fragments), reference materials reconstituted in simple assay buffers to act as calibration standards and QCs in LBAs remains a perennial problem (
To counteract the possibility of underestimating error during a typical fit-for-purpose biomarker method validation, a revised strategy based on the data here is proposed for future studies. The most important point in the revised strategy is the adoption of a confidence interval (2 s.d.; 95% confidence interval) favoured in diagnostic biomarker QC (
The stability of soluble protein biomarkers for analysis by ELISA assay is often assumed and studies of the effect of long-term storage of patient specimens before analysis (
In this study, stability was assessed in three different contexts: recombinant protein in a surrogate matrix (porcine plasma/serum); endogenous analytes in healthy volunteer plasma and endogenous analytes in cancer patients' plasma. Not surprisingly the greatest stability was observed in the surrogate matrix with recombinant proteins. In the healthy volunteers' plasma, a marked decrease in PDGF-BB concentration occurred after 8 months at −80°C, whereas an equally substantial increase in KGF occurred over the same time frame. Smaller changes were also recorded in VEGF-A, PlGF and FGFb, but these were closer to our acceptance limit of a 25% change for instability. Platelet-derived growth factor is a dimeric protein held together by two disulfide bonds, which are essential for correct folding and stability of the protein (
In cancer patients' plasma consistent increase in KGF concentration was also evident, even after 3 months of storage at −80°C, whereas a consistent decrease in PDGF-BB was recorded but only in platelet-depleted plasma. Sporadic changes (either increases or decreases in concentration) occurred with other analytes (VEGFR2, FGFb, HGF and VEGFR1) more randomly. These data would indicate that KGF is unstable in cancer patients' plasma and that PDGF-BB in the absence of platelets (where the majority of PDGF is normally located) is also unstable in plasma. The sporadic instability observed with other analytes may be caused by biological variables – such as disease stage, age or treatment regimen – additional to the duration of storage at −80°C. It should be noted, however, that different analytical platforms were used to conduct the stability studies. Single-plex ELISA was used with the recombinant proteins and multiplex ELISA was used for both healthy volunteer and cancer patients' samples. Therefore, some of the differences in stability profiles observed may be due to cross-platform variability.
In conducting stability studies, both statistical significance using the Wilcoxon signed-rank test and an increase or decrease greater than a predefined acceptance limit of 25% were required. This latter value was chosen as it is the default value for random error (imprecision) recommended in the fit-for-purpose approach to biomarker method validation (
Platelets are known to sequester a number of angiogenesis-regulatory proteins including FGFb, PDGF-BB, VEGF, VEGFR1, Ang-1, HGF and SDF-1
As most ELISAs are capable of only relative quantitation, one might expect different platforms, indeed even the same assay but sourced from different manufacturers, to yield discrepancies in the absolute concentrations measured in equivalent groups of patients (
In summary, the studies reported here have highlighted the need to conduct assay validation and to address sample handling issues, such as stability and the impact of platelet removal, before commencement of clinical trials if such biomarkers are to yield information useful for drug development and patient care.
Fit-for-purpose biomarker ELISA validation for use in clinical trials. Fit-for-purpose biomarker method validation was essentially a demonstration that a commercially available assay consistently performs within specification (either manufacturers or set in-house) using QCs before patient sample analysis and consisted of three stages. In the first stage the precision (as % CV) in the QCs was determined experimentally. In stage-2, a target CV acceptance limit was set against which the performance of future assays was evaluated. Stage-3 required that 2–3 additional assays fell within these target CVs for the QCs, to consider the assay valid for analysis of clinical trials samples. In the light of present data enhancement to this scheme is now recommended (see Discussion).
Stability of endogenous angiogenesis analytes in healthy-volunteer plasma. Plasma from healthy volunteers was analysed by multiplex ELISA (
Stability of endogenous angiogenesis analytes in both platelet-rich and platelet-depleted plasma in cancer patients. Plasma from five different colorectal cancer patients was analysed by multiplex ELISA before and after storage at −80°C for 3 months. The platelet-rich and platelet-depleted samples were produced as described in
Platelet count in plasma prepared using differing centrifugal techniques. Platelet counts were initially measured in WB collected from 20 colorectal cancer patients and in plasma (A) produced after centrifugation at 2000 g for 10 min. Plasma (A) was then subjected to re-centrifugation to produce three further samples: 2000 g for 15 min (A+B), 2000 g for 20 min (A+C) or 10 000 g for 20 min (A+D). The platelet count after spinning using method A+B was significantly (
Effect of platelet inclusion or removal and freeze–thaw on the concentration of angiogenesis analyte measured in cancer patients' plasma. (
Fit-for-purpose method validationa of 17 different ELISAs representing potential biomarkers of antivascular drugs using QC samples
|
|
|||||
|---|---|---|---|---|---|
|
|
|
|
|||
|
|
|
|
|
||
| CD105 | 4.83 | 4.53 | 8.58 | 20 | Yes |
| VEGF-A | 5.93 | 8.33 | 4.72 | 20 | Yes |
| PlGF | 5.68 | 6.07 | 5.91 | 20 | Yes |
| VEGF-receptor-1 (R1) | 10.1 | 12.3 | 10.9 | 20 | Yes |
| VEGF-receptor-2 (R2) | 5.29 | 3.15 | 4.21 | 20 | Yes |
| Angiopoetin-1 (Ang-1) | 2.67 | 2.98 | 2.25 | 20 | Yes |
| Angiopoetin-2 (Ang-2) | 7.73 | 3.17 | 10.4 | 20 | Yes |
| TIE-2 | 4.77 | 5.04 | 6.47 | 20 | Yes |
| VEGF-D | 14.0 | 13.1 | 7.08 | 20 | Yes |
| PDGF-BB | 11.1 | 8.86 | 10.6 | 20 | Yes |
| FGFb | 3.30 | 3.53 | 3.75 | 20 | Yes |
| IL-8 | 16.2 | 16.5 | 6.62 | 20 | Yes |
| SDF-1 |
4.40 | 4.96 | 6.25 | 20 | Yes |
| HGF | 15.8 | 9.39 | 7.10 | 20 | Yes |
| Osteopontin (OPN) | 6.91 | 7.52 | 6.83 | 20 | Yes |
| KGF | 17.6 | 11.0 | 5.00 | 20 | No |
| VEGF-C | 11.7 | 14.4 | 15.8 | 20 | No |
Abbreviations: CV=coefficient of variation; FGF=fibroblast growth factor; HGF=hepatocyte growth factor; IL=interleukin; KGF=keratinocyte growth factor; PDGF=platelet-derived growth factor; QC=quality control; VEGF=vascular endothelial growth factor.
Fit-for-purpose assay validation was conducted as described in
CV was calculated as a percentage using the following formula: the standard deviation in the 16 replicates divided by the mean value of the 16 replicates, multiplied by 100.
Duration of stabilitya of recombinant standards of angiogenesis biomarkers spiked in porcine plasma (P) and serum (S) and stored at different temperatures
|
|
||||||||
|---|---|---|---|---|---|---|---|---|
|
|
|
|||||||
|
|
|
|
|
|
|
|
|
|
| VEGF-A | 1 da | 1 d | 7 d | 1 d | 1 y | 1 y | 1 y | 1 y |
| PlGF | <1 d | <1 d | 7 d | <1 d | 1 m | 1 m | 1 y | 1 y |
| PDGF-BB | 1 d | 1 d | 1 d | 1 d | 1 y | 1 y | 1 y | 1 y |
| Ang-1 | 3 m | 3 m | ||||||
| Ang-2 | 3 m | 3 m | ||||||
| TIE-2 | 7 d | 7 d | 7 d | 7 d | 1 y | 1 y | 1 y | 1 y |
| FGF-b | 1 d | 1 d | 21 d | 21 d | 6 m | 6 m | 1 y | 1 y |
| SDF-1 |
7 d | 1 d | 7 d | 7 d | 1 m | 1 m | 1 m | 1 m |
| IL-8 | 7 d | 7 d | 7 d | 7 d | 1 m | 3 m | 3 m | 3 m |
| OPN | 1 d | 1 d | 7 d | 7 d | 3 m | 3 m | 1 y | 1 y |
| CD105 | 9 m | 9 m | ||||||
Abbreviations: Ang=angiopoetin; FGF=fibroblast growth factor; IL=interleukin; KGF=keratinocyte growth factor; OPN=osteopontin; PDGF=platelet-derived growth factor; VEGF=vascular endothelial growth factor.
d=day/s; m=month/s; y=year/s. Recombinant proteins were spiked at a mid-range concentration and stored at room temperature, 4°C, −20°C or −80°C for different durations of time up to 12 months in the case of −80°C. At defined intervals samples were retrieved for analysis by singleplex with instability being defined as a significant reduction in concentration (⩾25%) occurring between two time points.