Malaria rapid diagnostic tests (RDTs) are kits that generally include 20 to 25 test strips or cassettes, but only a single buffer vial. In field settings, laboratory staff occasionally uses saline, distilled water (liquids for parenteral drugs dilution) or tap water as substitutes for the RDT kit's buffer to compensate for the loss of a diluent bottle. The present study assessed the effect of buffer substitution on the RDT results.
Twenty-seven RDT brands were run with EDTA-blood samples of five malaria-free subjects, who were negative for rheumatoid factor and antinuclear antibodies. Saline, distilled water and tap water were used as substitute liquids. RDTs were also run with distilled water, without adding blood. Results were compared to those obtained with the RDT kit's buffer and
Only eight cassettes (in four RDT brands) showed no control line and were considered invalid. Visible test lines occurred for at least one malaria-free sample and one of the substitutes in 20/27 (74%) RDT brands (saline: n = 16; distilled water: n = 17; and tap water: n = 20), and in 15 RDTs which were run with distilled water only. They occurred for all
Replacement of the RDT kit's dedicated buffer by saline, distilled water and tap water can cause false-positive test results.
Malaria rapid diagnostic tests (RDTs) detect
RDTs are currently rolled out by national malaria control programmes in endemic settings as a tool for parasite based diagnosis in the scope of artemisinin-based combination therapy (ACT) [
Most malaria RDTs are marketed as kits that include material for 20 to 25 tests, such as lancets for finger pricking, test strips (available as dipstick, cassette or card box formats), transfer devices (pipettes, straws, capillaries or loops) and the buffer. Cassettes are usually individually packaged, and the number of lancets and transfer devices match the number of cassettes. All RDTs need a buffer to lyse the blood and to allow capillary flow along the nitrocellulose strip. Mostly, this buffer is supplied in a single bottle or dropper vial.
During field visits (particularly in Africa, ITM teams repeatedly observe availability problems of buffer vial: for instance, some cassettes were sent for testing in the ward, but the buffer vial did not return. To compensate, laboratory technicians took either a buffer vial from another kit (sometimes a kit of another brand), or used saline, distilled water (liquids for parenteral drugs dilution) and occasionally tap water as substitute liquids. Apparently, this substitution for the buffer seemed not to cause too much interference, as in all observed cases, there was enough background clearance and both control line and test lines were clearly distinguishable.
This phenomenon was also noted during a practical teaching session at the Institute of Tropical Medicine (ITM): students and staff were astonished to observe a visible HRP-2 line when the blood of one of the present authors (PG) was tested with a two-band RDT which was run with distilled water. PG had no symptoms of malaria, nor did he suffer from malaria in the recent past. When performing the RDT with the kit's dedicated buffer, there was no HRP-2 line visible. The HRP-2 line appeared upon retesting with distilled water and also when using saline and tap water as substitute liquids. As false-positive test line seemed to be the explanation, it was decided to explore this phenomenon.
EDTA-blood samples from five healthy subjects with no recent history of malaria were used. For all samples, the diagnosis of malaria was excluded by microscopy and species-specific PCR as previously described [
Malaria RDTs marketed as cassettes and folded card box were selected, and RDT brands commonly used in field settings were included. CE marking and FDA approval of the RDTs were recorded, as well as their presence on the World Health Organization (WHO) lists of RDT manufacturers and distributors complying with ISO13485:2003 or US FDA 21 CFR 820 production norms and their evaluation by the World Health Organization/Foundation for Innovative New Diagnostics (WHO/FIND) [
All RDT kits were used within their expiry date and had been stored at room temperature (maximum 25°C) before analysis. RDTs were assessed in the same run with the five subjects' blood and the following buffers: RDT kit's dedicated buffer, distilled water (Denolin, Brussels, Belgium), saline (NaCl 0.9%, Qualiphar, Bornem Belgium) and tap water from local supply. For each RDT and substitute liquid, the blood samples of all five subjects were used except when a particular RDT kit was finished. In addition, tests were run in the absence of blood, with distilled water as the substitute liquid.
All tests were performed according to the instructions of the manufacturer, except that samples were loaded with a pipette (Finnpipette, Helsinki, Finland) instead of the transfer device supplied by the manufacturer. Readings were performed by three readers at daylight assisted by a standard electric bulb, and within and not beyond the prescribed delay after application of the sample and buffer.
In case the control line did not appear, the result was considered as invalid and the test was repeated. RDT test lines were interpreted according to the manufacturers' instructions. In addition, test line intensities were scored into five categories: none (no line visible), faint (barely visible line), weak (paler than the control line), medium (equal to the control line) or strong (stronger than the control line) [
The appearance, shape and crispness of the control and test lines and the clearance of the background were compared with those obtained with the RDT kit's dedicated buffer and the two
Inter-reader reliability was assessed and expressed as percentage agreements for all three readers and kappa values for each pair of readers. To assess test reproducibility, a sample from one subject was tested upon five occasions for all RDT brands and the three substitute liquids. Two RDT brands were not included in the reproducibility assessment because of shortage of tests.
The study was approved by the Institutional Review Board of ITM and by the Ethical Committee of Antwerp University, Belgium.
Thirty different RDT brands were selected. As two brands from the same manufacturer showed bad clearance of the background upon testing with
Numbers of RDT brands showing false positive test lines when run with blood of malaria free subjects and substitute liquids
| RDT format | Numbers of different RDT brands assessed | Numbers of RDTs brands showing false positive test lines when run with substitute liquid* | |||
|---|---|---|---|---|---|
|
|
|||||
| Saline | Distilled water | Tap water | Any substitute liquid | ||
| Two band | 6 | 3 (1) | 3 (1) | 4 (1) | 4 (1) |
| Three band | 17 | 12 (9) | 11 (10) | 12 (9) | 12 (10) |
| Four band | 4 | 1 (0) | 3 (2) | 4 (2) | 4 (2) |
| Total | 27 | 16 (10) | 17 (13) | 20 (12) | 20 (13) |
*Numbers refer to the different RDT brands for which false positive test lines were visible in at least one sample. Between brackets: numbers of RDT brands for which false positive lines were visible in the samples of at least three subjects.
Numbers of false-positive test lines for the different
| RDT target antigen | Numbers of RDT brands (n = 27) detecting the target | Total numbers of tests performed with each substitute* | Numbers of false positive test lines when run with substitute liquid | ||
|---|---|---|---|---|---|
|
|
|||||
| Saline | Distilled water | Tap water | |||
| HRP-2 | 21 | 98 | 22 | 28 | 36 |
| Pf-pLDH | 5 | 25 | 8 | 9 | 9 |
| Pv-pLDH | 9 | 45 | 12 | 21 | 20 |
| Pan-pLDH | 14 | 70 | 30 | 25 | 33 |
| Aldolase | 3 | 11 | 1 | 5 | 5 |
*Each RDT brand was assessed with blood of five malaria-free subjects except one HRP-2-based kit and one HRP-2/aldolase-based kit that were assessed with two and one samples respectively.
When run with their kit's buffer, none of the RDT brands tested positive with any of the samples of the five malaria free subjects. Likewise, when tested with the two
When tested with a substitute liquid, there were eight invalid test results; they neither showed a control line upon repetition. Five of them occurred in a single RDT brand when assessed with distilled water. False-positive test lines were visible for at least one sample and substitute liquid in 20/27 (74%) RDT brands, at the following frequencies: saline: n = 16, distilled water n = 17 and tap water: n = 20 (Table
For distilled water and tap water as the substitute liquid, the RDT strips with false-positive lines showed a clearance of background similar to those observed with the RDT kit's buffer. In the case of saline, the background was less clear. Overall, the colour and crispness of the control and false-positive test lines were similar to those obtained with the RDT kits' buffers and
When tested with distilled water in the absence of blood, there was only a single invalid cassette, in which neither a control line appeared upon retesting. Control lines were observed in all remaining runs. Twenty-two false positive test lines were observed in 15 (55.6%) RDT brands and occurred in all
Numbers of false-positive test lines for the different
| RDT target antigen | Numbers of RDT brands (n = 27) detecting the target | Numbers of false positive test lines | Numbers of false positive test lines according to test line intensity | |||
|---|---|---|---|---|---|---|
|
|
||||||
| Faint | Weak | Medium | Strong | |||
| HRP-2 | 21 | 11 | 2 | 5 | 3 | 1 |
| Pf-pLDH | 5 | 2 | 1 | 1 | 0 | 0 |
| Pv-pLDH | 9 | 1 | 1 | 0 | 0 | 0 |
| Pan-pLDH | 14 | 6 | 1 | 4 | 1 | 0 |
| Aldolase | 3 | 2 | 1 | 1 | 0 | 0 |
The inter-observer agreement for positive and negative readings was high, with 96.4% overall agreement between the three observers and kappa values between 0.94-0.98 for each pair out of three observers. Inter-observer agreements were similar for all three substitute liquids. Discrepancies between line intensity readings were limited to one category of line intensity (
At reproducibility testing, false positive test lines occurred in at least four out of five runs in approximately half of the
CE-mark and non-CE labelled RDT brands were equally affected by the buffer substitution (11 out of 20, versus four out of seven). In addition, there were 11/22 WHO-GMP listed RTD brands affected versus four out of five RDTs that were not listed, 12/22 FIND evaluated brands versus three out of seven, and five out of nine RDTs from the WHO procurement list affected versus 10/18. Although the false-positive test lines occurred randomly among the different RDT brands, there were certain RDT brands that were more affected than others.
This study shows that replacing malaria RDT kit's buffers with saline, distilled water and tap water resulted in false-positive test lines in the majority of brands assessed. All
Malaria RDTs detect
The present study has its limitations. First, there are currently more than 80 different RDT brands marketed [
The presence of the control line is the result of the binding of the conjugated mouse-raised detection antibody to a goat-raised anti-mouse capture antibody: irrespective of the correct buffer there will be nearly always a control line visible as long as migration has been achieved. For the false-positive test lines, the situation is different. Their scattered distributions within the five malaria-free subjects and their moderate consistency upon retesting suggest nonspecific reactions, such as binding of the negatively charged colloidal gold conjugate to the positively charged capture antibodies. As the false-positive lines also occurred with distilled water in the absence of blood, a crucial role for the buffer is apparent. Apart from lysis of the red blood cells and allowing capillary migration of the sample along the strip, the buffer has other functions: it re-solubilizes blocking agents, such as detergents, polymers and proteins on the sample pad and the dried detection antibody-colloidal gold conjugate on the conjugate pad. Further, it ensures optimal pH and ionic strengths for the antigen-antibody reactions [
The high amount of false positive test lines with substitute liquids was unexpected and has, to the best of our knowledge, not yet been reported. The fact that the false positive test lines showed colour and crispness similar to those generated by
Although the phenomenon of buffer substitution was observed in various places and on different occasions, the real extent of this phenomenon in the field is unknown. Studies assessing errors made by end-users do not mention it [
False negative and false positive might delay/or exclude true diagnosis, with consequences that can go up to death for patients. The consequences of false-positive RDTs extend those of individual patient care and the non-justified prescription of ACT treatment: like other errors by end-users, poor performance of RDTs will erode the health care workers' confidence in RDT test results thereby hindering the implementation of RDTs in treatment algorithms and malaria control programs [
Prevention of buffer substitution can be addressed in several ways. From daily use and evaluations of RDTs, it is clear that some manufacturers already supply buffer vials with a volume in excess to the numbers of tests included. A solution could be the provision of more than one vial per kit, ideally (if costs are not too high), small plastic vials dedicated for each individual cassettes, as is already the case for individually wrapped packages for self-testing (example: CareStart® Malaria, Single Kit, Access Bio Inc., New Jersey, USA).
Complementary, RDT package inserts should mention not to use any other liquids apart from the buffer supplied with the kit. Likewise, the generic job aids on malaria RDT designed by WHO [
The issue of buffer substitution should further be addressed in RDT instructions and trainings at all levels of health care organization. With respect to the organization of RDT performance by the end users among health care workers, it is recommended that use of the correct buffer should be supervised by a laboratory officer and that the RDT kits' content should not be split.
In conclusion, buffer substitution in malaria RDTs causes false positive test lines in the majority of brands tested. Preventive measures in terms of product design, packaging, instructions manuals and trainings are needed to alert for this potential error.
ACT: Atimisinin based combination therapy; CE: Conformité Européenne; EDTA: Ethylene diamine tetra-acetic acid; FDA: Food and drug administration; FIND: Foundation for Innovative New Diagnostics; GMP: Good manufacturing practice; HRP-2: Histidine-rich protein 2; ISO: International organization for standardization; ITM: Institute of Tropical Medicine; NaCl: Sodium chloride;
The authors declare that they have no competing interests.
PG and JJ designed the study protocol, MM and JVDE made substantial contributions to the concept and design of the study. PG and MM carried out the test evaluations. PG, MM and JJ analyzed and interpret the results and drafted the manuscript. All authors critically reviewed the manuscript and approved the final manuscript.